Deviation Handling
Scenario: A technician finds a sample was left in the incubator longer than the method allows; the deviation runs from discovery to approval.
| # | Step | Owner | What happens |
|---|---|---|---|
| 1 | Record immediately | Technician | Open a deviation entry before leaving the bench: method, sample ID, what the procedure required, what happened, time found, and who was told. Do not report the sample result yet. |
| 2 | Contain | Laboratory supervisor | Hold the sample result and any other samples from the same run. Check whether the incubator timer or the handover caused the delay and whether other runs were exposed. |
| 3 | Assess impact and classify | Quality reviewer | Use the classification table. Determine whether the extra time could change the result and whether the method has validated limits that cover it. Classify as minor, major or critical with written reasoning. |
| 4 | Investigate cause | Laboratory supervisor | For major and above, find the cause: unclear shift handover, missing timer alarm, or a misread instruction. Propose corrections to the procedure or equipment. |
| 5 | Decide disposition | Quality manager | Approve one outcome: result valid with justification, retest, or invalidate. Sign and date. Informing the client, where relevant, follows the complaint and notification procedures. |
| 6 | Close and trend | Quality reviewer | Close once actions are complete. Add the deviation to the monthly trend table by method and cause; three on the same method open a CAPA. |